Jarrett-Isotech Water Triple Point Cells

The Water Triple Point is the most important fixed point, the only point common to the ITS-90 and the Thermodynamic Temperature Scale. It is an essential temperature reference standard for every temperature laboratory. The Jarrett-Isotech water triple point cell is the best standard. Not all cells are the same. Accept no inferior device.

  • Fifty Years of Proven Use
  • Uncertainty 0.000070°C
  • Quartz Glass and Isotopic Analysis Available
MODEL TYPES PERFORMANCE
Premium Water Triple Point Cells – Borosilicate Glass These Premium Cells can be calibrated to an uncertainty of +/- 0.000070°C (0.07mK) A11-50-270*, A13-50-270*, B11-50-270, B11-65-270*, B13-65-270, B16-65-270 *also available in Quartz Glass.
Reproducibility The equilibrium temperature of a cell will repeat to within ±0.000,02°C of the mean equilibrium temperature.
Stability After equilibrium is reached, the mean temperature of the inner melt of an ice mantle will remain constant to within ±0.000,01°C for as long as the mantle can be preserved (up to 90 days in some instances).
Further Laboratory Water Triple Point Cells These Cells have shorter immersion depth and can be calibrated to an uncertainty of ±0.000100°C (0.1mK). B12-40-210, B12-46-210.
Life Cells made from Borosilicate Glass may drift lower in temperature by up to 0.1mK after 10 to 20 years. Cells made from Quartz Glass are expected to last 10 times longer.
Accuracy The equilibrium of temperature of the Jarrett-Isotech Triple Point of Water Cell is within ±20μK of the 2005 definition as described in CCT/05-07/rev(2). Isotopic Analysis can be provided to give the exact composition within ±3μK.
Slim Water Triple Point Cells B8-30-130 This small cell fits into portable Dry Block Calibrators for use outside of the calibration industry. It can be calibrated to ±0.0005°C (0.5mK).