首頁  >  廠牌分類  >  ISOTECH 一級與二級實驗室  >  二級標準低溫校正設備  >  18205 液態氮沸點爐 -185°C to -196°C

 

Isotech Nitrogen Boiling Point Apparatus is designed specifically to realise and maintain the liquid/vapour equilibrium (boiling point) of nitrogen or argon, for the calibration of thermometers on the International Temperature Scale of 1990.

  • Boiling Point of Liquid Nitrogen
  • Gas Manifold for Thermometers
  • Self Contained - Bench Mounted

 

DESCRIPTION

Isotech Nitrogen Boiling Point Apparatus is designed specifically to realise and maintain the liquid/vapour equilibrium (boiling point) of nitrogen or argon, for the calibration of thermometers on the International Temperature Scale of 1990.

The Boiling Point Apparatus is self-contained and refrigerated by liquid nitrogen or liquid argon, which must be supplied by the user. Either liquid is suitable for the purpose. Liquid nitrogen is generally less costly and more readily available than liquid argon. The Apparatus does not require electrical power for its operation.

The Boiling Point Apparatus will maintain the liquid-vapour equilibrium of nitrogen (-195.798°C) or of argon (-185.8468°C) indefinitely, provided boiled-off gas is replenished.

ITS-90 specifies the triple point of argon (-189.3442°C) as the low end of the long-stem Standard Platinum Resistance Thermometer range. As a practical matter, realisation of this triple point is a costly (in equipment and time) and complicated process.

Most laboratories will choose to calibrate this end of the platinum range by comparison of the thermometer under test with a thermometer of known calibration. National Laboratories themselves will invariably calibrate thermometers submitted to them by comparison, realising the actual argon triple point only infrequently for calibration of their own reference thermometers. The National Physical Laboratory of England makes this statement:

“Most thermometers (submitted for calibration) will involve measurements (by) … comparison with NPL standards in a bath of liquid nitrogen (about -196°C)”

Model 18205 Comparator is designed for precisely such comparison calibration. It comprises a stainless steel dewar, an inner equalising block having wells for 3 thermometers, top connections for filling and monitoring the level of liquid coolant, a pressure safety blow-off and a manifold which may be used to thermally tie the thermometers under test to the equalising block with helium gas (optional).

Since the slopes (dR/dT) of Standard Platinum Resistance Thermometers are very similar at any temperature, calibration uncertainties not larger than 0.002K can be obtained at a small fraction of the cost of an absolute calibration.

 

 

TECHNICAL FEATURE VALUE
Model ITL-M-18205
Temperature Range -185°C or -196°C nominal
Uncertainty ±0.002°C – the temperature distribution across the copper block is less than 2mK. Uncertainty must be added issued with the calibration certificate from the National Laboratory. Extra uncertainties will exist if dissimilar probes are compared.
Power N/A
Dimensions Overall Diameter 142mm, Overall Height 570mm, Weight 14kgs
How to order ITL-M-18205 Nitrogen Boiling Point Apparatus. Note – the Nitrogen Boiling Point Apparatus may be adapted for use with liquid argon. Please contact the factory for details.
18205 型錄
PDF

相關產品

This model is a lower cost alternative to the ITL M 18205. This model is a simple apparatus open to the atmosphere compr
The Isotech Mercury Cell Model 17724 is constructed in a rugged, sealed stainless steel enclosure allowing the triple po
Second only to the Water Triple Point and in many ways, because of its ease of use and purity, superior to it, is the Ga
The Isotech Argon Triple Point Apparatus is a robust, simple to use and affordable solution for the realisation of the a